Actual Comprehensive Real Exam Questions And Verified
Answers ( 100% Correct Answers) Already Graded A+ |
Guaranteed Success | Newest EXAM | Just Released!!
Gas specific, whether the outlet/inlet is threaded or is a non-
interchangeable
quick coupler. 5.1.5.1 page 45 - ANSWER-On each station
outlet/inlet for
medical gases or vacuums shall
be_______?
8 inches, and PS quarter inch 3/8" inch OD .3 inches 5.1.5.11
page 45 - ANSWER-For station outlets and inlets,factory install
copper INLETS tubes on station outlets extending no further
than _________from the body of the terminal shall not be less
than _______ size, with _________ minimum inside diameter.
8 inches, 3/8" inch NPS 1/2"OD, .4 inches 5.1.5.12 page 45 -
ANSWER-Factory
installed copper outlet tubes on station inlet extending
no further than
_________ from the body of the terminal shall not be less than
_______ size
with ______ minimum inside
diameter.
,#1 They shall not be interchangeable with other systems
including medical surgical vacuum.
#2 components necessary for maintenance of a WAGD
specificity shall be legibly marked to identify them as
components of a WAGD inlet. #3 they shall be of a type
appropriate for the flow and vacuum the required by the
facilities gas and aesthetic machine.
#4 They shall be located to avoid physical damage to the end.
5.1.5.16 page 46 - ANSWER-WAGD networks shower provide a
WAGD in like in all locations were nitrous oxide or halogenated
anesthetic gas is intended to be administered. Station in that's
for a WAGD service shall have the following additional
characteristics.
+ or -5% 5.1.8.1.4 page 47 - ANSWER-On pressure and vacuum
indicators, the accuracy of a digital indicator shall be _______
of operating pressure at which they are used.
35°F, -22°F 5.1.9.2.4 —10. —— 12 page 48 - ANSWER-On a
MASTER
ALARM panel for medical gas and vacuum systems shout
each include the
following signals. Medical air dew high alarm for each
compressor site to indicate when the line pressure dewpoint is
greater than ______. An instrument air dewpoint hi alarm for
each compressor site to indicate when line pressure dewpoint
is greater than ______.
, Area alarm 20% 5.1.9.4.2 page 49 - ANSWER-AREA ALARM
panels for medical gas systems shall indicate if pressure in the
lines in the area being monitored increases or decreases by
______ percent from the normal line pressure.
Air compressor systems, medical surgical vacuum pump
systems, WAGD systems, instrument air systems, and
proportioning systems. 5.1.9.5 page 49 - ANSWER-Local alarm
shall be installed to monitor the functions of what systems?
Medical air dewpoint 35°
instrument air dewpoint -22°
high carbon level 10 ppm's
or higher
Proportioning systems 19.5 to23.5% oxygen range.
Medical air/WAGD reserve under the man if pump ceases to
operate
5.1.9.5.4 page 50 - ANSWER-The local alarm will activate when
the following will happens?
Plugged or capped my manufacturer 5.1.10.1.2 page 50 -
ANSWER-All the
tubes fittings, valves, and other components shall be
_________ by the
___________and sealed until prepared for
installation
3 feet 5.1.10.1.6 page 50 - ANSWER-CMT (corrugated medical
tubing) piping