NFPA 99 MedGas for 6010 Installer ||Verified
Exam!!||Newest With Complete All 300 Questions And
Correct Detailed Answers|| Newest Exam!!!
The locations within 6' of the patient's bed and 7'-6" from
the ceiling, where examinations take place? - Answer-
patient care vicinity
Where liquid ring compressors for medical air are used,
the air receiver shall have a local alarm which will shut
down the system due to? - Answer-high water level
An area where failure of a system is not likely to cause
any physical impact on a patient. - Answer-Category 4
Space
Building or structure, used to house hyperbaric chambers
for medical applications and procedures at pressures
above normal atmospheric pressure. - Answer-hyperbaric
facility
,2|Page
The warning system that provides a continuous visible and
audible vigilant watch on Category 1 & Category 2 piping
systems. - Answer-area alarm
Moisture concentration of the outlet testing during the
verifier purity test shall not exceed how many ppm? -
Answer-500
This alarm shall be triggered once the backup instrument
air compressor is energized. - Answer-local alarm
The components which make up surface mounted medical
gas rails shall have a melting point of no less than _____.
- Answer-1000 degrees F
When the carbon monoxide level in a medical air system
exceeds this ppm level, the local alarm is triggered. -
Answer-10
The most remote outlet & the source gas undergo the
verifiers piping purity test. The ppm non-methane variation
between the two shall not exceed? - Answer-5 ppm
,3|Page
Building used on a 24-hour basis for medical, psychiatric,
or surgical care of 4 or more patients. - Answer-hospital
The ppm gaseous hydrocarbons in the medical air system
shall be less than? - Answer-25 ppm
At any level of demand, the medical air dryers shall
provide air with a dew point below this temperature? -
Answer-32 degrees F
Which of the following is NOT permitted to have the
system supply source located outside?
A. vacuum pumps
B. manifolds for gas cylinders
C. instrument air compressors
D. medical air compressors - Answer-medical air
compressors
Regulators, filters, and dryers shall be sized to meet the
peak calculated demand. The piping format for these items
shall be at a minimum? - Answer-duplexed
, 4|Page
They can allow an existing system, which is not in
compliance with the NFPA 99 Code to remain in service. -
Answer-Authority Having Jurisdiction (AHJ)
Medical air supply systems operating at different
pressures shall be provided with? - Answer-pressure
regulator
The removal of the outlet connection on a cryogenic liquid
container will cause the container to become? - Answer-
unusable
A building, space, room, or group of rooms intended to
serve activities involving procedures for investigation,
diagnosis, or treatment in which flammable, combustible,
or oxidizing materials are to be used. - Answer-laboratory
Carbon dioxide and nitrous oxide cylinders shall be kept
from reaching temperatures below _____. - Answer-20
degrees F