CA Fluoro tests Flash Cards with correct answers 100% An image intensifiers photocathode causes ___ to be given off in direct proportion to the intensity of the fluorescent light: a. remnant radiation b. electrons c. x-rays d. light photons - Correct Answer electrons What fluoroscopic television system device decodes the video signal into a visible image? a. Vidicon camera b. Camera control unit c. Television monitor d. Electron gun - Correct Answer Television Monitor Which of the following materials are used in the construction of a fluoroscopic tabletop? 1. Lead 2. Aluminum 3. Bakelite 4. Carbon fiber - Correct Answer 2, 3, and 4 Geometric factors in the shape of the image intensification tube b. Focal-film-distance (FFD) c. Object-to-image distance (OID) d. Minification gain - Correct Answer ... If the entrance skin dose is measured from the surface of the tabletop, the x-ray source is located: a. either over or under the table b. under the table c. over the table d. not enough information provided - Correct Answer b. Under the table The mottle level of an image can be adjusted by: 1. changing mA 2. changing kVp 3. changing the exposure factors 4. increasing the number of available x-ray photons - Correct Answer All of the above During a fluoroscopic exam, the main source of scattered radiation is produced from: a. the patient b. x-ray source c. table top d. leakage radiation - Correct Answer the patient Excessive lighting in the fluoroscopy room: a. Helps to perceive a dim image as brighter b. Helps to properly adjust the overall luminance of the TV image c. Enhances visualization of a black and white TV image d. May result in increasing the technical operating factors - Correct Answer May result in increasing the technical operating factors Suitable gonad shields must be used: a. Whenever gonads are within the useful x-raybeam b. When shielding of gonads does not interfere with the diagnosis c. On all potentially procreative patients d. At all times - Correct Answer When shielding of gonads does not interfere with the diagnosis Which of the following are true concerning cine film? 1. Available film sizes are: 70 mm, 100 mm, and 110 mm 2. The larger the film size the better the image quality 3. Nearly all cine-fluorography exams involve studies of the GI system 4. Delivers approximately the same dose as routine photospot image recording - Correct Answer 2 only The exposure rate at the tabletop may not exceed five rads per minute except: 1. when using the ABC on large patients 2. during magnification mode 3. where higher exposure rates are allowed - Correct Answer b. 1, 2 and 3 only Images produced with photo spot film cameras ___ spot films taken with conventional cassettes. 1. use a higher mA than 2. require about 1/2 to 1/3 the dose of 3. exhibit superior image quality compared to - Correct Answer require about 1/2 to 1/3 the dose The ability of an imaging system to differentiate small objects that are closely spaced as separate objects is referred to as: a. Differentiation b. Resolution c. Statistical fluctuation d. Detector contrast - Correct Answer b. Resolution The photoconductive target in a television camera pick-up tube: a. Conducts electrons across to the target b. Allows a proportional electron charge to be generated at a particular position c. Creates a video signal d. Creates an electronic current which is converted into a video image . - Correct Answer Allows a proportional electron charge to be generated at a particular position Size distortion is primarily caused by: a. Focal-film-distance (FFD) b. Geometric factors in the shape of the image intensification tube c. Minification gain d. Object-to-image distance (OID) - Correct Answer Object-to-image distance (OID)

Show more Read less
Institution
California Fluoroscopy
Course
California Fluoroscopy

Content preview

An image intensifiers photocathode causes ___ to be given off in direct proportion to the intensity of the
fluorescent light:

a. remnant radiation

b. electrons

c. x-rays

d. light photons - Correct Answer electrons



What fluoroscopic television system device decodes the video signal into a visible image?

a. Vidicon camera

b. Camera control unit

c. Television monitor

d. Electron gun - Correct Answer Television Monitor



Which of the following materials are used in the construction of a fluoroscopic tabletop?

1. Lead

2. Aluminum

3. Bakelite

4. Carbon fiber - Correct Answer 2, 3, and 4



Geometric factors in the shape of the image intensification tube

b. Focal-film-distance (FFD)

c. Object-to-image distance (OID)

d. Minification gain - Correct Answer ...



If the entrance skin dose is measured from the surface of the tabletop, the x-ray source is located:

a. either over or under the table

b. under the table

,c. over the table

d. not enough information provided - Correct Answer b. Under the table



The mottle level of an image can be adjusted by:

1. changing mA

2. changing kVp

3. changing the exposure factors

4. increasing the number of available x-ray photons - Correct Answer All of the above



During a fluoroscopic exam, the main source of scattered radiation is produced from:

a. the patient

b. x-ray source

c. table top

d. leakage radiation - Correct Answer the patient



Excessive lighting in the fluoroscopy room:

a. Helps to perceive a dim image as brighter

b. Helps to properly adjust the overall luminance of the TV image

c. Enhances visualization of a black and white TV image

d. May result in increasing the technical operating factors - Correct Answer May result in increasing the
technical operating factors



Suitable gonad shields must be used:

a. Whenever gonads are within the useful x-raybeam

b. When shielding of gonads does not interfere with the diagnosis

c. On all potentially procreative patients

d. At all times - Correct Answer When shielding of gonads does not interfere with the diagnosis



Which of the following are true concerning cine film?

,1. Available film sizes are: 70 mm, 100 mm, and 110 mm

2. The larger the film size the better the image quality

3. Nearly all cine-fluorography exams involve studies of the GI system

4. Delivers approximately the same dose as routine photospot image recording - Correct Answer 2 only



The exposure rate at the tabletop may not exceed five rads per minute except:

1. when using the ABC on large patients

2. during magnification mode

3. where higher exposure rates are allowed - Correct Answer b. 1, 2 and 3 only



Images produced with photo spot film cameras ___ spot films taken with conventional cassettes.

1. use a higher mA than

2. require about 1/2 to 1/3 the dose of

3. exhibit superior image quality compared to - Correct Answer require about 1/2 to 1/3 the dose



The ability of an imaging system to differentiate small objects that are closely spaced as separate objects
is referred to as:

a. Differentiation

b. Resolution

c. Statistical fluctuation

d. Detector contrast - Correct Answer b. Resolution



The photoconductive target in a television camera pick-up tube:

a. Conducts electrons across to the target

b. Allows a proportional electron charge to be generated at a particular position

c. Creates a video signal

d. Creates an electronic current which is converted into a video image



. - Correct Answer Allows a proportional electron charge to be generated at a particular position

, Size distortion is primarily caused by:

a. Focal-film-distance (FFD)

b. Geometric factors in the shape of the image intensification tube

c. Minification gain

d. Object-to-image distance (OID) - Correct Answer Object-to-image distance (OID)



The overall luminescence of the TV image may be properly adjusted by:

1. increasing room lighting

2. adjusting the contrast control button on the TV monitor

3. adjusting the brightness control button on the TV monitor - Correct Answer 2 and 3 only



Which of the following statements is true concerning fluoroscopic grids?

a. Reduces scattered radiation produced in the image intensifier

b. The typical grid ratio is 12:1

c. High ratio grids are preferred

d. Located between the patient and input layer of the image intensifier - Correct Answer Located
between the patient and input layer of the image intensifier



One of the most important actions a fluoroscopist can take to avoid unnecessary radiation exposure to
the patient is to use appropriate:

a. Illumination

b. Primary beam filtration

c. Collimation

d. Protective lead curtains - Correct Answer Collimation



The increase in effective brightness due to the size reduction of the output image size is referred to as:

a. Total gain

b. Minification gain

Written for

Institution
California Fluoroscopy
Course
California Fluoroscopy

Document information

Uploaded on
March 22, 2026
Number of pages
42
Written in
2025/2026
Type
Exam (elaborations)
Contains
Questions & answers

Subjects

$10.49
Get access to the full document:

Wrong document? Swap it for free Within 14 days of purchase and before downloading, you can choose a different document. You can simply spend the amount again.
Written by students who passed
Immediately available after payment
Read online or as PDF

Get to know the seller

Seller avatar
Reputation scores are based on the amount of documents a seller has sold for a fee and the reviews they have received for those documents. There are three levels: Bronze, Silver and Gold. The better the reputation, the more your can rely on the quality of the sellers work.
EXAMSSITE West Virgina University
Follow You need to be logged in order to follow users or courses
Sold
16
Member since
2 year
Number of followers
3
Documents
2053
Last sold
2 months ago
EXAMS CORNER LAW,LITERATURE ,NURSING

Welcome to my Stuvia study hub — your ultimate destination for high-quality, reliable, and exam-focused study resources designed to help you succeed with confidence. If you’re looking for well-structured notes, expertly prepared summaries, and exam materials that actually make a difference in your grades, you’re in the right place. My goal is simple: to save you time, reduce your stress, and help you achieve top results without feeling overwhelmed. I understand how challenging academic life can be. Between lectures, assignments, deadlines, and personal commitments, it often feels like there aren’t enough hours in the day. That’s exactly why I created this collection of study materials — to give you a smarter, more efficient way to learn. Every document I upload is created with attention to detail, clarity, and effectiveness. These are not rushed or generic notes. They are carefully organized, easy to follow, and tailored to match real course content and exam expectations. I focus on delivering materials that simplify complex topics and highlight what truly matters for your exams. Start exploring the notes today and take a step closer to achieving the results you deserve.

Read more Read less
3.0

1 reviews

5
0
4
0
3
1
2
0
1
0

Recently viewed by you

Why students choose Stuvia

Created by fellow students, verified by reviews

Quality you can trust: written by students who passed their tests and reviewed by others who've used these notes.

Didn't get what you expected? Choose another document

No worries! You can instantly pick a different document that better fits what you're looking for.

Pay as you like, start learning right away

No subscription, no commitments. Pay the way you're used to via credit card and download your PDF document instantly.

Student with book image

“Bought, downloaded, and aced it. It really can be that simple.”

Alisha Student

Working on your references?

Create accurate citations in APA, MLA and Harvard with our free citation generator.

Working on your references?

Frequently asked questions