Exam Coverage:
Medical gas system installation, brazing and piping practices, safety
procedures, code compliance, system testing, and documentation.
The Med Gas Installer 6010 exam evaluates a candidate's ability to
properly install medical gas and vacuum systems in healthcare
settings. It emphasizes correct installation techniques, safe brazing,
and adherence to standards from the National Fire Protection
Association, particularly NFPA 99, ensuring systems are safe,
compliant, and reliable.
,Mains & branches in medical gas systems shall be no less than ____
1/2 inch
Drops to station outlets and inlets shall be no less than ____ NPS
1/2 inch
In-line check valves shall have threaded purge points of ____ NPT
1/8 inch
,Copper outlet tubes on station inlets shall be no less than ____ O.D.
1/2 inch
Connecting tubing for gauges and alarm devices shall be permitted to
be ____ NPS
1/4 inch
Category 1 Dental
Facilities utilizing General anesthesia & Deep sedation
, Instrument air standby headers shall be sufficient for ____ of normal
operation from attached cylinder
1 hour
Reserves cylinders for Liquid Ring compressors shall be sufficient for
____ normal operation, when installed
1-hour
Quality of medical air shall not have permanent particulates sized ____
or larger in the air
1 Micron