●● Liquid containers shall have additional product identification visible
from all directions with a minimum of _______ inch high letters such as
a 360-degree wrap-a-round tape for medical liquid containers. Answer: 2
●● Which systems shall not be permitted to be located in the same
indoor enclosure? Answer: Bulk cryogenic liquid system
●● Cylinder and cryogenic liquid container outlet connections shall be
designed as to render the attachment point _______________ when
removal of the outlet connection is attempted. Answer: Unusable
●● Cylinders in use and in storage shall be prevented from reaching
temperatures in excess of _________________. Answer: 125 degrees F
●● What is required of the system if another alarm condition occurs
while the audible alarm is manually silenced? Answer: The audible
signal will beep
●● Mechanical exhaust for central supply locations shall not be more
than __________. Answer: 500 cfm
,●● All pressure relief valves shall be made of brass, bronze,
_________________. Answer: Stainless Steel
●● Metallic flexible joints shall be permitted in the pipeline where
required for expansion joints, seismic protection, thermal expansion, or
vibration control, Shall have the following _____________. Answer: -
Suitable for service at 300 psig at or above 1000 degrees F
- Provided with brazing extensions
- Made of Bronze, Copper, or Stainless Steel for all wet surfaces.
(ALL THE ABOVE)
●● The pressure in the main line of each medical gas system, the
operation and condition of the source of supply and the reverse source
(if any) shall be monitored at the ___________________ alarm Answer:
Master
●● Manifolds for gas cylinders without reverse shall consist of
_____________ equal headers; each with an average day's supply, but
not fewer than ____________ connections. Answer: Two - Two
●● Markings used for the identification of contents of cylinders shall not
be __________ or ___________, including decals, tags, and stenciled
marks, except those labels, tags used for indication cylinder status.
Answer: Defaced or removed
, (Both A & B)
●● Manifolds for cryogenic liquid containers shall consist of ________
equal headers, each having sufficient number of liquid containers for an
average day's supply. Answer: 2
●● The government body of the facility or its designee shall establish
the following areas. What area is anticipated use for patient care?
Answer: - Category 1 spaces
- Category 2 spaces
- Category 3 spaces
(ALL THE ABOVE)
●● Manifold for cryogenic liquid containers shall indicate when or at a
predetermined set point before the reserve contents fall to one day's
average supply, indication _____________ Answer: Reserve low
●● Installer of micro-bulk cryogenic liquid systems shall be qualified to
meet __________ Standard, Guide for Medical Gas Installations at the
Consumers Sites. Answer: CGA M-1
●● When is the initial pressure test conducted by the installer? Answer:
After blowdown of the distrubtion piping