Which of the following is true?
A. transition S1 to S0 is phosphorescence and the transition T1 to S0 is fluorescence, both radiational
B. transition s1 to s0 is fluorescence and the transition T1 to s0 is phosphorescence, both radiational
C. transition s1 to s0 is fluorescence and the transition T1 to s1 is phosphorescence, both are radtiational
D. none of the above - Correct Answers B. transition s1 to s0 is fluorescence and the transition T1 to s0 is
phosphorescence, both radiational
The purpose of the blank is to
A. cancel readings due to the color of the reagents
B. account and subtract the absorbency due to imperfections in the cuvette
C. cancel and subtract readings due to the absorbance of the reagents added to the analyte
D. none of the above - Correct AnswersC. cancel and subtract readings due to the absorbance of the
reagents added to the analyte
A _____ lamp is used as a source of light in UV spectrophotometer, where as _____ lamp is used in case
of atomic absorption spectroscopy
A. deuterium arc, hollow cathode
B. deuterium arc, tungsten
C. hollow cathode, tungsten
D. tungsten, hollow cathode - Correct AnswersA. deuterium arc, hollow cathode
The purpose of the ashing step in graphite-furnace atomic absorption spectroscopy is to:
A. vaporize metal atoms to be analyzed
B. cause the metal to form carbides which are more volatile
C. convert the graphite furnace to ash to avoid carbide formation
, D. burn off organic species that would interfere with elemental analysis - Correct AnswersD. burn off
organic species that would interfere with elemental analysis
How should the concentration of a colorless sample be determined?
A. using a UV spectrophotometer
B. using a VIS spectrophotometer
C. Either a UV or VIS spectrophotometer
D. using an indicator solution and pH paper - Correct AnswersA. using a UV spectrophotometer
Self absorption can be expected in ____ at ______ analyte concentrations
A. spectrophotometry, high
B. flourometry, high
C. atomic absorption, low
D. flourometry, low - Correct AnswersB. flourometry, high
The purpose of the flame in the ______ spectroscopy is to ______
A. atomic absorption, purify the sample
B. atomic absorption, desolvate and atomize the analyte atoms in a sample
C. ICP, excite the analyte atoms
D. atomic absorption, ionize the analyte atoms - Correct AnswersB. atomic absorption, desolvate and
atomize the analyte atoms in a sample
Light is passed through a solution with a light path of 2.0 cm and the absorbance is recorded as 0.40 . If
light is passed through the same solution as the same wavelength and the light path is reduced to 1.0
cm the absorbance should be :
A. 0.80
B. 0.40
C. 0.20
D. 0.10 - Correct AnswersC. 0.20