& Answers (Grade A+)
The ACR does not need to see or require information for QC tests
that are specific to DBT, but must be included in the report to the
FDA.
Select one:
True
False -
correct answer ✅true
Before imaging with DBT, facilities must meet all MQSA applicable
requirements. These requirements include _____________.
Select one:
a. The unit must undergo a MEE prior to use
b. Personnel must obtain 8 hours of DBT training
c. Facility must follow manufacturer's recommended QC procedures
d. All of the above -
correct answer ✅all of the above
There is not a QC standard in place for all DBT systems.
,Mammography DBT 2 Exam Questions
& Answers (Grade A+)
Select one:
True
False -
correct answer ✅true
In order to perform DBT QC tests you must follow the equipment
vendors QC manual.
Select one:
True
False -
correct answer ✅true
Medical Physicists test performed on the Hologic Selenia DBT
system include ________ and ______. -
correct answer ✅Phantom Image Quality, AEC Performance
Technologist Test performed on the DBT system include ______ and
_______. -
correct answer ✅Geometry Calibration, Artifact Evaluation
, Mammography DBT 2 Exam Questions
& Answers (Grade A+)
The artifact evaluation test in the 3D DBT Hologic System requires
the use of the LFS with an aluminum filter.
Select one:
True
False -
correct answer ✅true
The HVL QC test requires the technologists to use "Zero Degree
Tomo" in 3D mode, _______. yes or no. -
correct answer ✅yes
The performance criteria for the Dose QC test in the Hologic DBT
System for 2D + 3D is AGD<3.0 mGy.
Select one:
True
False -
correct answer ✅true