Questions & Answers (Grade A+)
The laboratory manager receives a complaint from the ICU about
turnaround times for coagulation tests. The first step in problem
solving should be -
correct answer ✅Gather data on current times by shift
Which action by the phlebotomist will comply with the CAP patient
safety goal "to improve patient and sample identification at the
time of specimen collection" and The Joint Commission Patient
Safety Goal to "improve the accuracy of patient identification"? -
correct answer ✅Match the name and medical record number on
the patient's ID bracelet to the nae and medical record number on
the preprinted collection label
If the information on a swab collected in the micro department is
different from the information on the requisition form, what do you
do -
correct answer ✅Document error and communicate to the lab
supervisor so that the risk manager conducts a root cause analysis
Preanalytical variables in laboratory testing includes -
correct answer ✅specimen acceptability
,ASCP Laboratory Operations Exam
Questions & Answers (Grade A+)
The first procedure to be followed if the blood gas instrument is
out-of-control for all parameters is: -
correct answer ✅recalibrate, then repeat control
A graph shows a positive correlation with points fairly close to the
line. The new method can be described as -
correct answer ✅good correlation with constant bias
A graph has 4 lines. A being the steepest and D being the least-
steep. The line that demonstrates a proportional error relationship
is -
correct answer ✅Line D
Shown above is a levy-jenningsQC chart, which represents control
values for 13 consecutive analyses for a particular serum
constituent. If the 14th value is below the -2SD limit, Which of the
following should be done (points 7-14 will be descreasing) -
correct answer ✅Analysis system should be checked for a
deteriorating component
The precision of an instrument is validated by -
correct answer ✅running the same sample multiple times
,ASCP Laboratory Operations Exam
Questions & Answers (Grade A+)
The mean value of a series of HGB controls was found to be 15.2
g/dL, and the standard deviation was calculated at 0.20. Acceptable
control range for the laboratory is +/- 2 SD. Which of the following
represents the allowable limits for the control? -
correct answer ✅14.8-15.6 g/dL
In QC, +/- 2 SD from the mean includes what % of the sample
population -
correct answer ✅95%
Upon completion of a run of cholesterol tests, the technician
recognizes that the controls are not within the 2 SD confidence
range. What s the appropriate course of action -
correct answer ✅Run a new set of controls and repeat specimens
The following data were calculated on a series of 30 determination
of serum uric acid control:
mean=5.8mg/dL
1SD =0.15 mg/dL
If the confidence limits are set at 2SD, what is the allowable range -
correct answer ✅5.50-6.10 mg/dL
, ASCP Laboratory Operations Exam
Questions & Answers (Grade A+)
An index of precision is statistically known as -
correct answer ✅Coefficient of variation (CV)
the lower CV= better precision
The term used to describe reproducibility is -
correct answer ✅precision
The ability of a procedure to measure only the component(s) it
claims to measure is called -
correct answer ✅specificity
The extent to which measurements agree with the true value of the
quality being measured is known as -
correct answer ✅accuracy
Diagnostic specificity is defined as the percentage of individual -
correct answer ✅without a given disease who have a negative
result by a given test
If the correlation coefficient (r) of two variables is 0: -
correct answer ✅There is an absence of correlation