If the output power setting is at 50% and this indicates 0dB, increasing the power by: Correct Answer: 3dB will increase the power output by a factor of 2 An increase in power of 3dB indicates an increase of 2X the original power. Every 3dB that the power increases will double the power output. 50% x 2 = 100% The doppler tracing demonstrates electrical noise. How can you reduce or eliminate this artifact? Correct Answer: Increase the wall filter settings Noise can be reduced by decreasing doppler gain, increasing wall filter settings or trying another wall outlet. Which of the following is true regarding the Curie point of a piezoelectric element? Correct Answer: If the active PE crystal is heated to this temperature, the element will lose its piezoelectric properties. Certain ceramic elements can be polarized by heating them above their Curie point. If the active element is heated again, it will lose its piezoelectric properties What dimension of the mass is being measured on this image? Correct Answer: AP The spatial resolution on a CRT monitor is directly related to: Correct Answer: The number of lines per frame The vacuum tube TV (CRT) monitor interfaces two fields together to from an image frame. The more lines in each field, the better the spatial resolution on the image What system component is not needed when an LCD monitor is used? Correct Answer: digital to analog converter LCD monitors accept the digital signal and do not need the component The matching layer for a txd is usually made of plastic and its approx: Correct Answer: ¼ of the probe wavelength in thickness If the radius of a blood vessel decreases by 50% what happens to the flow volume? Correct Answer: decreases by 16 Radius decreases by 1/2, flow volume decreases by 2 to the fourth power (16x); if the radius decreases to ⅓ original size, the flow volume decreases by 3 to the fourth power (81x). Note the significant change in flow value from, a 50% reduction in radius to a 66% reduction in radius Why is there no color flow displa

Show more Read less
Institution
Course

Content preview

ULTRASOUND PHYSICS REVIEW| 113 PAGES| 837 QUESTIONS| WITH
COMPLETE SOLUTIONS
If the output power setting is at 50% and this indicates 0dB, increasing the power by: Correct Answer:
3dB will increase the power output by a factor of 2

An increase in power of 3dB indicates an increase of 2X the original power. Every 3dB that the power
increases will double the power output. 50% x 2 = 100%

The doppler tracing demonstrates electrical noise. How can you reduce or eliminate this artifact?
Correct Answer: Increase the wall filter settings

Noise can be reduced by decreasing doppler gain, increasing wall filter settings or trying another wall
outlet.

Which of the following is true regarding the Curie point of a piezoelectric element? Correct Answer: If
the active PE crystal is heated to this temperature, the element will lose its piezoelectric properties.

Certain ceramic elements can be polarized by heating them above their Curie point. If the active
element is heated again, it will lose its piezoelectric properties

What dimension of the mass is being measured on this image? Correct Answer: AP

The spatial resolution on a CRT monitor is directly related to: Correct Answer: The number of lines per
frame

The vacuum tube TV (CRT) monitor interfaces two fields together to from an image frame. The more
lines in each field, the better the spatial resolution on the image

What system component is not needed when an LCD monitor is used? Correct Answer: digital to analog
converter

LCD monitors accept the digital signal and do not need the component

The matching layer for a txd is usually made of plastic and its approx: Correct Answer: ¼ of the probe
wavelength in thickness

If the radius of a blood vessel decreases by 50% what happens to the flow volume? Correct Answer:
decreases by 16

Radius decreases by 1/2, flow volume decreases by 2 to the fourth power (16x); if the radius decreases
to ⅓ original size, the flow volume decreases by 3 to the fourth power (81x). Note the significant change
in flow value from, a 50% reduction in radius to a 66% reduction in radius

Why is there no color flow displayed in the right hepatic vein: Correct Answer: The blood flow is
perpendicular to the beam at the zero degree angle and undetectable

,The txd is producing a beam that is perpendicular to he blood flow. The txd is evaluating the emotion of
reflectors moving toward or away from it and assigning the appropriate color. The blood in the vessel is
moving left to right, therefore no motion to and from the probe is detected or displayed as color on the
image

You switch from a 5MHz curved probe to a 2.5 MHz sector probe to image deeper in the abdomen. How
is the wavelength affected by the change to the new probe? Correct Answer: Doubles

Wavelength and freq are inversely related. If the freq is reduced by 50%, the wavelength will double

Curved PE element and acoustic lenses will improve which two types of resolution? Correct Answer:
elevational (vertical) and lateral

Curved elements and lenses will improve focusing and decreases slice thickness. This will improve lateral
and slice thickness resolution

Elevational resolution is determined by the height of the ultrasound beam. It is the ability to distinguish
two objects close together along the y-axis (vertical = elevational). The shorter = thinner the beam the
better the elevational resolution

Lateral resolution is defined as the ability of the system to distinguish two points in the direction
perpendicular to the direction of the ultrasound beam. It is also known as azimuthal resolution. Lateral
resolution is affected by the width of the beam and the depth of imaging

Which of the following imaging techniques would be used to best evaluate the volume of an irregular
mass? Correct Answer: 3D rendering

3D rendering refers to the creation of an image based on data acquired during the exam. It is considered
a type of post-processing. 3D techniques more accurately measure the volume of structures than 2D
imaging.

If you are performing an abdominal exam but inadvertently select the carotid preset on the machine,
how will this affect your exam? Correct Answer: Depth settings will be very superficial causing potential
non-visualization of the abdominal structures

The carotid preset will be set to a very low imaging depth due to the superficial course of the carotid
arteries. The preset for abdomen is set to begin at a much higher depth setting.

Which of the following is NOT adjustable by the Sonographer? Correct Answer: Demodulation

Demodulation occurs as the signal is converted into positive voltages (rectification) and smoothed
across the peaks of the voltage pulses. This process assists in processing the voltages for proper image
display of the returned signals. This is a preset function of the US system and not operator adjustable.

For this reason the process is often called "demodulation" or "detection". Though coined during the
development of radio technology, they sound intuitively correct for ultrasound as well. The first
ultrasound signal processing technique is quadrature detection, or IQ as it's sometimes called

,_________________ refers to the ability of a system to detect low-amplitude echoes accurately. Correct
Answer: sensitivity

Sensitivity refers to the ability of a system to detect low-amplitude echoes accurately.
The bandwidth refers to the range of frequencies produced by the probe.
The dynamic range refers to the number of levels of gray displayed on the image.
The quality factor refers to the purity of the sound wave.

What is the best way to increase beam penetration? Correct Answer: Decrease frequency

Decreasing transducer frequency will increase sound penetration

________________ is defined as the range of frequencies contained in the US pulse. Correct Answer:
Bandwidth

Decreasing the pulse repetition period leads to: Correct Answer: a decrease in the maximum depth that
can be imaged

If the PRP decreases, the PRF will increase which means shorter pulses are emitted more quickly. The
shorter pulses will travel a shorter distance than the pulses emitted from a lower PRF. The imaging
depth will decrease with a decrease in PRP.

Pulse repetition frequency (PRF): Number of pulses occurring in 1 s. Usually expressed in kHz. Pulse
repetition period (PRP): Time from the beginning of one pulse to the beginning of the next. Usually
expressed in microseconds

The more bits per pixel, the _________________ shades of grey on the image and ________________
improves. Correct Answer: more, contrast resolution

What part of the ultrasound system is responsible for changing the element function back and forth
from transmit to receive? Correct Answer: beam former

The beam former is responsible for switching the element function back and forth between transmit and
receive.

Beamforming is a common signal processing technique used to create directional or spatial selectivity of
signals sent to or received from an array of sensors or antennae. These arrays can be found in many
different devices that transmit and receive either electromagnetic or acoustic waves.

If the Sonographer increases the size of the field of view, the frame rate will ___________________.
Correct Answer: decrease

increasing the area to be scanned will require an increase in time to process the increased # of scan lines
for display causing a decreased frame rate. Frame rate is indirectly related to the size of the field of
view. Increased size of the FOV = slower (decreased) frame rate. If the question read "If the size of the
field of view doubled, how would the frame rate be affected?" Then the answer could be more specific
to say that the frame rate would decrease by 50%.

, In order to improve spatial resolution and the signal-to-noise ratio, ______________________ uses
frequency modulation to manipulate the transmitted beam. Correct Answer: Coded excitation

Spatial resolution is a measure of the accuracy or detail of a graphic display, expressed as dots per inch,
pixels per line, lines per millimeter, etc. It is a measure of how fine an image is, usually expressed in dots
per inch (dpi).

As the transducer frequency increases, the spatial pulse length ____________________. Correct
Answer: decreases

Spatial pulse length is the length of one pulse composed of multiple cycles. SPL = # cycles X wavelength;
Normal US range 0.1 - 1.0mm; The shorter the pulse length, the better the axial resolution. Axial
resolution = 1/2 SPL

If you increase the sector angle from 45 to 90 degrees and reduce the line density by 50%, what
happens to the frame rate? Correct Answer: unchanged

Doubling the sector size will double the number of scan lines and reduce the frame rate by 50%. BUT if
the line density is reduced by 50% when the sector size doubles, the frame rate does not change.
Producing and processing more scan lines takes for time = frame rate is reduced. # and density of scan
lines are indirectly related to frame rate. Because the sector size was doubled, the frame rate will be
reduced by 1/2 UNLESS line density changes too.

Which operator control can be used to adjust the image contrast? Correct Answer: dynamic range

Dynamic Range/Compression - adjusts the range of grays displayed on the image; changes the contrast
level of the image; increased/larger dynamic range = decreased image contrast

The pulse repetition period in 2D imaging is determined by: Correct Answer: imaging depth

The imaging depth and PRF are directly related. PRP = 1 / PRF The imaging depth establishes the PRF and
PRP.

The range equation is used to determine the: Correct Answer: distance to a reflector

The range equation is used by the US system to place the echoes on the image properly.

When Harmonic imaging is turned on, it primarily improves ___________ resolution. Correct Answer:
Lateral

Harmonic imaging improves spatial resolution with the greatest effect on lateral resolution. Harmonic
beams are narrower than the fundamental beam so lateral resolution is improved. Other benefits of
harmonic imaging include reduced image clutter and increased signal to noise ratio.

What portion of the TGC curve is the deepest region attenuation compensation can occur? Correct
Answer: knee

Written for

Institution
Course

Document information

Uploaded on
October 26, 2022
Number of pages
113
Written in
2022/2023
Type
Exam (elaborations)
Contains
Questions & answers

Subjects

$15.99
Get access to the full document:

Wrong document? Swap it for free Within 14 days of purchase and before downloading, you can choose a different document. You can simply spend the amount again.
Written by students who passed
Immediately available after payment
Read online or as PDF


Also available in package deal

Get to know the seller

Seller avatar
Reputation scores are based on the amount of documents a seller has sold for a fee and the reviews they have received for those documents. There are three levels: Bronze, Silver and Gold. The better the reputation, the more your can rely on the quality of the sellers work.
Classroom NURSING
Follow You need to be logged in order to follow users or courses
Sold
4885
Member since
4 year
Number of followers
3233
Documents
55440
Last sold
2 days ago
NURSING

Assignments, Case Studies, Research, Essay writing service, Questions and Answers, Discussions etc. for students who want to see results twice as fast. I have done papers of various topics and complexities. I am punctual and always submit work on-deadline. I write engaging and informative content on all subjects. Send me your research papers, case studies, psychology papers, etc, and I’ll do them to the best of my abilities. Writing is my passion when it comes to academic work. I’ve got a good sense of structure and enjoy finding interesting ways to deliver information in any given paper. I love impressing clients with my work, and I am very punctual about deadlines. Send me your assignment and I’ll take it to the next level. I strive for my content to be of the highest quality. Your wishes come first— send me your requirements and I’ll make a piece of work with fresh ideas, consistent structure, and following the academic formatting rules. For every student you refer to me with an order that is completed and paid transparently, I will do one assignment for you, free of charge!!!!!!!!!!!!

Read more Read less
4.0

1192 reviews

5
631
4
216
3
196
2
40
1
109

Recently viewed by you

Why students choose Stuvia

Created by fellow students, verified by reviews

Quality you can trust: written by students who passed their tests and reviewed by others who've used these notes.

Didn't get what you expected? Choose another document

No worries! You can instantly pick a different document that better fits what you're looking for.

Pay as you like, start learning right away

No subscription, no commitments. Pay the way you're used to via credit card and download your PDF document instantly.

Student with book image

“Bought, downloaded, and aced it. It really can be that simple.”

Alisha Student

Working on your references?

Create accurate citations in APA, MLA and Harvard with our free citation generator.

Working on your references?

Frequently asked questions