CCHT WATER
CHLORINE TESTING - answer REVERSE OSMOSIS (RO) MACHINES RUNS 15
MINUTES BEFORE PERFORMING ANY TOTAL CHLORINE TESTING
100 ML - answer THE CORRECT AMOUNT OF WATER TO COLLECT FOR TOTAL
CHLORINE TESTING
RPC ultra low total chlorine strips - answer STRIPS USED FOR CHLORINE
TESTING
HOW LONG ARE RPC STRIPS - answer IMMERSED IN SAMPLE WATER FOR 60
SECONDS
WHEN SHOULD CHLORINE BE TESTED - answer PRIOR TO TREATMENTS IN
THE MORNING AND EVERY 3-4 HOURS WHILE PATIENTS ARE DIALYZING
RPC ULTRA -LOW TOTAL CHLORINE STRIPS TEST - answer CHLORINE IN THE
WATER AFTER CARBON TANK AND POST RO
0.00 to 0.09 ppm - answer total chlorine levels that are patient safe AND
DOCUMENT IN "TMS" OR PAPER LOG
FIRST SAMPLE - answer SHOULD BE TAKEN FROM POST CARBON TANK 1
(WORKER )
SECOND SAMPLE - answer POST RO
IF CHLORINE IS NOTED POST RO - answer DO NOT INITIATE ANY PATIENT
TREATMENT, AND DOCUMENT PER FACILITY POLICY AND PROCEDURE
CHLORINE BREAKTHROUGH - answer OCCURS POST CARBON TANK 1
(WORKER), IMMEDIATE ADDITIONAL TESTING SHOULD TAKE PLACE AFTER
CARBON TANK 2 (POLISHER)
CARBON TANK 2 (POLISHER) - answer IF TANK SHOWS RESULTS ABOVE THE
ACTION LEVEL IMMEDIATELY STOP TREATMENTS
BYPASS - answer YOU WILL PUT PATIENTS IN THIS IF CARBON TANK 2
(POLISHER) TEST POSTIVE FOR CHLORINE , DOCUMENT TMS OR TCL- 2 LOG
EVERY 2 HOURS
CHLORINE TESTING - answer REVERSE OSMOSIS (RO) MACHINES RUNS 15
MINUTES BEFORE PERFORMING ANY TOTAL CHLORINE TESTING
100 ML - answer THE CORRECT AMOUNT OF WATER TO COLLECT FOR TOTAL
CHLORINE TESTING
RPC ultra low total chlorine strips - answer STRIPS USED FOR CHLORINE
TESTING
HOW LONG ARE RPC STRIPS - answer IMMERSED IN SAMPLE WATER FOR 60
SECONDS
WHEN SHOULD CHLORINE BE TESTED - answer PRIOR TO TREATMENTS IN
THE MORNING AND EVERY 3-4 HOURS WHILE PATIENTS ARE DIALYZING
RPC ULTRA -LOW TOTAL CHLORINE STRIPS TEST - answer CHLORINE IN THE
WATER AFTER CARBON TANK AND POST RO
0.00 to 0.09 ppm - answer total chlorine levels that are patient safe AND
DOCUMENT IN "TMS" OR PAPER LOG
FIRST SAMPLE - answer SHOULD BE TAKEN FROM POST CARBON TANK 1
(WORKER )
SECOND SAMPLE - answer POST RO
IF CHLORINE IS NOTED POST RO - answer DO NOT INITIATE ANY PATIENT
TREATMENT, AND DOCUMENT PER FACILITY POLICY AND PROCEDURE
CHLORINE BREAKTHROUGH - answer OCCURS POST CARBON TANK 1
(WORKER), IMMEDIATE ADDITIONAL TESTING SHOULD TAKE PLACE AFTER
CARBON TANK 2 (POLISHER)
CARBON TANK 2 (POLISHER) - answer IF TANK SHOWS RESULTS ABOVE THE
ACTION LEVEL IMMEDIATELY STOP TREATMENTS
BYPASS - answer YOU WILL PUT PATIENTS IN THIS IF CARBON TANK 2
(POLISHER) TEST POSTIVE FOR CHLORINE , DOCUMENT TMS OR TCL- 2 LOG
EVERY 2 HOURS