Which gain will increase the risk of bioeffects?
a. transmit
b. receive
c. gain
d. overall gain - ANSWER a. transmit
Which of the following is considered the memory of an ultrasound instrument?
a. pulser
b. transducer
c. receiver
d. beam former
e. digital scan converter
f. display monitor - ANSWER e. digital scan converter
Spectral broadening suggest:
a. laminar flow
b. turbulent flow
c. flow away from the transducer
d. flow towards the transducer - ANSWER b. turbulent flow
With pulse wave Doppler axial resolution: - ANSWER degrades
What color is usually use in color flow Doppler variance map to indicate turbulent
flow? - ANSWER green
Doppler high pass filter eliminates: - ANSWER high amplitude signals; low velocity
flow
The Doppler wall filter is considered a: - ANSWER high pass filter
The Nyquist Limit is equal to:
a. propagation speed x frequency
b. propagation speed/ wavelength
c. PRF/2
d. 4xV^2 - ANSWER c. PRF/2
Multiple echoes, equally spaced originating from a gas bubble is called:
a. comet tail
b. ring down
c. slice thickness artifact
, d. multipath - ANSWER b. ring down
An increase in red blood cell velocity will_____ the Doppler Shift:
a. increase
b. decrease
c. not change
d. cannot be predicted - ANSWER a. increase
The primary advantage of CW Doppler is:
a. range resolution
b. range discrimination
c. range ambiguity
d. range gating
e. absence of sampling rate - ANSWER e. absence of sampling rate
The range for ultrasound begins:
a. 20 Hz
b. 20,000 Hz
c. 2 MHz
d. 15 MHz - ANSWER b. 20,000 Hz (useful frequency range for clinical imaging is
2MHz-10MHz)
Which instrument control affects the amplitude (dB) of the outgoing signal:
a. gain
b. receiver gain
c. TGC
d. transmit gain
e. overall gain - ANSWER d. transmit gain; amplitude can be affected by the
sonographer
Which of the following is used to convert Doppler shift information into color:
a. high PRF
b. fast Fourier transform
c. autocorrelation
d. time interval histogram - ANSWER c. autocorrelation
b is incorrect because fast Fourier transform is the method used to process
conventional Doppler shift information but is too slow a method for color Doppler
Pulse repetition frequency - ANSWER Number of pulses that an ultrasound system
transmits into the body each second. Hz. Typically 4-15 KHz. PRF and depth of a
view are inversely related: When system is imaging deeper, the pulse repetition
frequency is lower.
Also the number of pulses created each second
The best way to increase the near field length of an ultrasound beam is to: -
ANSWER increase transducer diameter
The range of frequencies produced by a damped PZT element is called:
a. fundamental frequencies
b. Q Factor
2024 VERSION B ACTUAL QUESTIONS AND
CORRECT VERIFIED ANSWERS |GUARANTEED
PASS
Which gain will increase the risk of bioeffects?
a. transmit
b. receive
c. gain
d. overall gain - ANSWER a. transmit
Which of the following is considered the memory of an ultrasound instrument?
a. pulser
b. transducer
c. receiver
d. beam former
e. digital scan converter
f. display monitor - ANSWER e. digital scan converter
Spectral broadening suggest:
a. laminar flow
b. turbulent flow
c. flow away from the transducer
d. flow towards the transducer - ANSWER b. turbulent flow
With pulse wave Doppler axial resolution: - ANSWER degrades
What color is usually use in color flow Doppler variance map to indicate turbulent
flow? - ANSWER green
Doppler high pass filter eliminates: - ANSWER high amplitude signals; low velocity
flow
The Doppler wall filter is considered a: - ANSWER high pass filter
The Nyquist Limit is equal to:
a. propagation speed x frequency
b. propagation speed/ wavelength
c. PRF/2
d. 4xV^2 - ANSWER c. PRF/2
Multiple echoes, equally spaced originating from a gas bubble is called:
a. comet tail
b. ring down
c. slice thickness artifact
, d. multipath - ANSWER b. ring down
An increase in red blood cell velocity will_____ the Doppler Shift:
a. increase
b. decrease
c. not change
d. cannot be predicted - ANSWER a. increase
The primary advantage of CW Doppler is:
a. range resolution
b. range discrimination
c. range ambiguity
d. range gating
e. absence of sampling rate - ANSWER e. absence of sampling rate
The range for ultrasound begins:
a. 20 Hz
b. 20,000 Hz
c. 2 MHz
d. 15 MHz - ANSWER b. 20,000 Hz (useful frequency range for clinical imaging is
2MHz-10MHz)
Which instrument control affects the amplitude (dB) of the outgoing signal:
a. gain
b. receiver gain
c. TGC
d. transmit gain
e. overall gain - ANSWER d. transmit gain; amplitude can be affected by the
sonographer
Which of the following is used to convert Doppler shift information into color:
a. high PRF
b. fast Fourier transform
c. autocorrelation
d. time interval histogram - ANSWER c. autocorrelation
b is incorrect because fast Fourier transform is the method used to process
conventional Doppler shift information but is too slow a method for color Doppler
Pulse repetition frequency - ANSWER Number of pulses that an ultrasound system
transmits into the body each second. Hz. Typically 4-15 KHz. PRF and depth of a
view are inversely related: When system is imaging deeper, the pulse repetition
frequency is lower.
Also the number of pulses created each second
The best way to increase the near field length of an ultrasound beam is to: -
ANSWER increase transducer diameter
The range of frequencies produced by a damped PZT element is called:
a. fundamental frequencies
b. Q Factor