CCHT WATER EXAM QUESTIONS AND ANSWERS WITH COMPLETE
SOLUTIONS VERIFIED LATEST UPDATE
Practice questions for this set
Learn 1/7 Study using Learn
AFTER CHLORINE BREAKTHROUGH , WHO MUST APPROVE TO RESUME NORMAL
OPERATION AND SIGN INTO TMS OR TCL -2 LOG
Select the correct term
CLINICAL AND TECHNICAL
1EMS SUPERVISOR
2MANAGEMENT
3MEDICAL DIRECTOR
4MIXING BICARBONATE FOR DIALYSIS
Don't know?
Terms in this set (37)
REVERSE OSMOSIS (RO) MACHINES RUNS 15 MINUTES BEFORE PERFORMING ANY
CHLORINE TESTING
TOTAL CHLORINE TESTING
100 ML THE CORRECT AMOUNT OF WATER TO COLLECT FOR TOTAL CHLORINE TESTING
RPC ultra low total chlorine strips STRIPS USED FOR CHLORINE TESTING
HOW LONG ARE RPC STRIPS IMMERSED IN SAMPLE WATER FOR 60 SECONDS
PRIOR TO TREATMENTS IN THE MORNING AND EVERY 3-4 HOURS WHILE
WHEN SHOULD CHLORINE BE TESTED
PATIENTS ARE DIALYZING
RPC ULTRA -LOW TOTAL CHLORINE CHLORINE IN THE WATER AFTER CARBON TANK AND POST RO
STRIPS TEST
0.00 to 0.09 ppm total chlorine levels that are patient safe AND DOCUMENT IN "TMS" OR PAPER LOG
FIRST SAMPLE SHOULD BE TAKEN FROM POST CARBON TANK 1 (WORKER )
SECOND SAMPLE POST RO
DO NOT INITIATE ANY PATIENT TREATMENT, AND DOCUMENT PER FACILITY
IF CHLORINE IS NOTED POST RO
POLICY AND PROCEDURE
OCCURS POST CARBON TANK 1 (WORKER), IMMEDIATE ADDITIONAL TESTING
CHLORINE BREAKTHROUGH
SHOULD TAKE PLACE AFTER CARBON TANK 2 (POLISHER)
IF TANK SHOWS RESULTS ABOVE THE ACTION LEVEL IMMEDIATELY STOP
CARBON TANK 2 (POLISHER)
TREATMENTS