CERTIFIED QUESTIONS AND Answers
Verified 100% Correct
The air kerma rate must not exceed ___ from scattered radiation of radioscopic
equipment - CORRECT ANSWER- 2 mGy/h
HVL of Aluminum for given tube voltages? (70kV-150kV) - CORRECT ANSWER- 70:
2.5mm
80: 2.9mm
90: 3.2mm
100: 3.6mm
110: 3.9mm
120: 4.3mm
130: 4.7mm
140: 5.0mm
150: 5.4mm
How frequently should IP's and cassettes be cleaned and inspected? - CORRECT
ANSWER- Cleaned: monthly
Visual inspection: Weekly
What are daily QC tests for CR/DR/Fluoro/CT equipment - CORRECT ANSWER- -
Warm up (all)
-Meter operation (all)
-Equipment condition (all)
-System movements (fluoro)
-Visual assessment of electronic display devices (all)
What are weekly QC tests for CR/DR/Fluoro/CT equipment - CORRECT ANSWER- -
Visual inspection of cleanliness of system (all)
-Laser film printer operation (all)
-CT # accuracy
-CT noise
-CT uniformity
,What are monthly QC tests for CR/DR/Fluoro/CT equipment - CORRECT ANSWER- -
Cassette, screen, IP plate cleaning (CR)
-Retake analysis (CR/DR)
-Electronic display device performance (all)
-Laser film printer operation (all)
-CT slice thickness
-CT # calibration
-CT # linearity
What should the monthly retake analysis be below? - CORRECT ANSWER- 5% What
are quarterly QC tests for CR/DR/Fluoro/CT - CORRECT ANSWER- -Collimator
operation (CR/DR/Fluoro)
-Interlocks (all) (on doors to prevent exposure if door is open)
-Compression devices (Fluoro)
-Table angling/motion (Fluoro)
-Fluoro protection devices (lead drapes = .25mm of pb)
-CT patient support movement
-CT spatial resolution
-CT low contrast detectability
What are the annual QC tests for CR/DR/Fluoro/CT - CORRECT ANSWER- -Accuracy
of loading factors (CR/DR/fluoro)
-Output reproducibility
-Output linearity
-Beam filtration
-AEC
-Beam/light field alignment
- Beam collimation
-Grid performance
-EI
-Dynamic Range
-Spatial resolution
-Integrity of protective equipment
Flexion/ Extension views of the C spine assess? - CORRECT ANSWER-
Anteroposterior mobility
CR angle for an AP coccyx is - CORRECT ANSWER- 10 degrees caudad
What does ERCP stand for, what 2 things is it done for? - CORRECT ANSWER-
Endoscopic Retrograde Cholangiopancreatography
, 1. Diagnose biliary and pancreatic pathological conditions
2. Investigate gallbladder, CBD, pancreatic ducts, and intervention
Test tool used to test focal spot - CORRECT ANSWER- Slit camera, pinhole camera,
star pattern
State the line focus principle - CORRECT ANSWER- By angling the anode target the
effective focal spot is made much smaller than the actual focal spot
What are the effects of a smaller anode angle? (more vertical) - CORRECT ANSWER- -
Smaller effective focal spot
-Better spatial resolution
-More anode heel effect
-Smaller max size of collimation
What is the grid ratio? - CORRECT ANSWER- The height of the lead strips divided by
the distance between lead strips
Ratio = h/D
Explain bit conversions for grids - CORRECT ANSWER- 5:1 Grid: 1 bit
8:1 / air gap: 2 bits
12:1 grid: 2.5 bits
SID bit conversions - CORRECT ANSWER- 100cm->122cm: 0.5 bit
100cm->180cm: 1.5 bit
122cm-> 180cm: 1 bit
*every 25cm change is a 0.5 bit change*
Bit conversions for part thickness - CORRECT ANSWER- Every 5cm change = 1 bit
What errors will cause grid cut off with a parallel grid? - CORRECT ANSWER- -SID too
small
-Angling against grid lines
-Grid and CR not perpendicular
What errors will cause grid cut off with a focussed grid - CORRECT ANSWER- -
Incorrect SID (off focus)
-Angling against grid lines
- Off center
- Grid and CR not perpendicular