Which gain will increase the risk of bio effects?
a. transmit
b. receive
c. gain
d. overall gain
ANSWER: a. transmit
Which of the following is considered the memory of an ultrasound instrument?
a. pulser
b. transducer
c. receiver
d. beam former
e. digital scan converter
f. display monitor
ANSWER: e. digital scan converter
Spectral broadening suggest:
a. laminar flow
b. turbulent flow
, 2|Page
c. flow away from the transducer
d. flow towards the transducer
ANSWER: b. turbulent flow
With pulse wave Doppler axial resolution:
ANSWER: degrades
What color is usually use in color flow Doppler variance map to indicate turbulent
flow? ANSWER: green
Doppler high pass filter eliminates:
ANSWER: high amplitude signals; low velocity flow
The Nyquist Limit is equal to:
a. propagation speed x frequency
b. propagation speed/ wavelength
c. PRF/2
d. 4xV^2
ANSWER: c. PRF/2
Multiple echoes, equally spaced originating from a gas bubble is called: