CBSPD EXAM QUESTIONS AND ANSWERS WITH
COMPLETE SOLUTIONS VERIFIED LATEST RELEASE
2025/2026
Terms in this set (332)
Superheating can occur If the steam pressure is too low for the set temperature
Wet packs Wet steam can cause
68-73 degrees AAMI recommends that the temperature in the
sterilization area be:
30%-60% AAMI recommends that the humidity is the sterilization
area be
10 The recommended ventilation in the sterilization
are is—— air exchanges per hour under
positive pressure
75-85/10/negative AAMI recommends that the temperature in the
sterilizer access area be —- with — air
exchanges per hour and under — pressure
Decon, the sterilizer The only 3 areas in SPD that should have negative
access area, EVS pressure are:
closet
sterility — is the absence of all forms of microbial life, including
bacterial spores
An item is considered sterile if there is less than ——
chance that a viable
One in a million; 10(-6)
microorganism has survived the sterilization process.
1/
12
, 8/16/25, 10:47
AM
This is reflected numerically as
—-
Sterility Assurance Level SAL=
Steam — kills microorganisms by denature great the protein
steam sterilization — is the most common form of sterilization for
items that are heat- and moisture- tolerant
saturated steam under Three parameters of steam sterilization cycles are
pressure, time, and
temperature
Dynamic-air-removal — are used in prevacuum sterilized and in steam-flush
cycles pressure-pulse (SFPP)
Prevacuum sterilizers — remove air by drawing series of vacuums
with an injection of steam between each
vacuum pull.
Bowie-Dick test Performed daily to detect air leaks or
inadequate air removal from prevacuum
sterilzers
Gravity-displacement Incoming steam displaces residual air through
cycles the drain at the bottom of the sterilizer
chamber:
Gravity-displacement Because there is no mechanical air removal in
a —, a Bowie-Dick test is not required for — or
sterilizers
Boiler In most healthcare facilities, steam is generated inside of
a:
60-80psig The required pressure of steam as it enters the jacket is
usually:
28-30 lbs of pressure For a sterilizer operating at 270 degrees, the sterilizer
2/
12