Aftercoolers, where required, shall be provided with individual ___.
A. Condensate traps
B. Flexible connectors
C. Pressure indicators
D. Shutoff valves - Answer- A. Condensate traps
Carbon monoxide shall activate local alarms when the CO level exceeds ___.
A. 20 ppm
B. 10 ppm
C. 15 ppm
D. 5 ppm - Answer- B. 10 ppm
Compressors that are monitored for gaseous hydrocarbons shall be monitored on a ___
basis.
A. Quarterly
B. Annual
C. Daily
D. Monthly` - Answer- A. Quarterly
Dew point shall activate the required alarms when the dew point at system pressure
exceeds ___.
A. 35 F
B. 32 F
C. 39 F
D. 42 F - Answer- A. 35 F
Liquid ring compressor system shall have a reserve medical air standby header (1HR)
or a ___.
A. Manifold
B. Bulk system
C. Condensate traps
D. Compressor - Answer- D. Compressor
Liquid ring compressors, service water and seal water shall be treated to control ___.
A. Waterborne pathogens
B. Chlorine
C. Hyper chlorination
D. Both A & B - Answer- D. Both A & B
Medical air compressor systems shall have pressure relief valves set at ___ percent
above line pressure.
A. 30
B. 40