NFPA 99 6040 practice set Questions and Answers Verified 100% Correct The entrance to indoor locations containing positive pressure gases other than oxygen and medical air shall be labeled including: ______________ A Positive Pressure Gases B No Smoking or Open Flame C Room May Have Insufficient Oxygen - Open Door and Allow Room to Ventilate Before Entering D All of the above - ANSWER -All of the above If heat is necessary to keep an indoor manifold enclosure above 20oF and the manufacturer's recommended minimum ambient temperature it may be provided by: A Any in-room electric heating element B In-room heating element with a maximum temperature of 266oF C Fuel fired equipment in the room D Heat blanket with maximum temperature of 300oF - ANSWER -In-room heating element with a maximum temperature of 266oF Two entries/exits are required for ___________________ gas supplies. A Outdoor Cryogenic Gas Supplies only B Outdoor Cylinder Gas Supplies in an enclosure 200ft2 only C Indoor Gas Supplies only D Both A & B - ANSWER -Both A & B An existing outdoor Nitrous Oxide Manifold may continued to be used since it seldom falls below 20oF. A TRUE B FALSE - ANSWER -FALSE Indoor manifold rooms must be separated by ______ fire rated walls and floors with a ______ opening. A 1 hour, 1 hour B 2 hour, 1 hour C 1 hour, 3/4 hour D No wall fire rating is required - ANSWER -1 hour, 3/4 hour Only full or in-use cylinders are required to be prevented from falling by being secured with racks, chains, or other fastenings. Empty cylinders are not required to be secured. A TRUE B FALSE - ANSWER -FALSE Oxygen supplies greater than 350 psig may not use _______________ materials. A Copper B Stainless Steel C Polymeric D Brass - ANSWER -Polymeric Gas supply source relief valve vent lines must be labeled ____________. A with the proper pressure in psig B with medical gas pipe labels C by painting the correct CGA color D in a manner to distinguish them from medical gas pipeline - ANSWER -in a manner to distinguish them from medical gas pipeline Manifold rooms utilizing natural ventilation shall A have 2 non-closeable louvered openings B be vented directly to the outside without ductwork C provide cross ventilation D all of the above - ANSWER -all of the above Indoor manifold enclosures utilizing mechanical ventilation shall maintain positive pressure A TRUE B FALSE - ANSWER -FALSE A gas cylinder header consists of all of the following components with the exception of: A Filter B Pressure Indicator C Local alarm D Regulator - ANSWER -Local alarm A header for cryogenic liquid containers must have a ________________. A pressure relief valve B purge ports C quick disconnect D none of the above - ANSWER -pressure relief valve A __________ must be installed at the connection between the header and each cylinder lead (pigtail) to prevent a single pigtail leak from causing pressure loss of all cylinders connected to that same header. A Warning light B Local Alarm C Gauge D Check Valve - ANSWER -Check Valve Full and empty cylinder storage may not be in the same enclosure as a central supply system (medical or support gas manifold). A TRUE B FALSE - ANSWER -FALSE Permitted areas for piped medical gases outlets include all of the following with exception of: A Direct Respiration by a Patient B Calibration of patient respiration devices C Endoscopy or Sterile Processing Instrument Blow Down D Simulation centers for education of health care professionals - ANSWER Endoscopy or Sterile Processing Instrument Blow Down An auxiliary valve source connection must consist of all of the following except: A Installed on the source side (upstream) of the source valve B Must the same size as the main line C Include a removable plug or cap D Must be normally closed and secured - ANSWER -Installed on the source side (upstream) of the source valve Facilities may supply ______________ through a Concentrator Supply Source. A USP 93 Oxygen B 99% Oxygen Only C Compressed Medical Air D Nitrous Oxide - ANSWER -USP 93 Oxygen The primary and secondary headers for a gas cylinder manifold must each have an equal number of cylinders. A TRUE

Meer zien Lees minder
Instelling
NFPA 99 6040
Vak
NFPA 99 6040

Voorbeeld van de inhoud

NFPA 99 6040 practice set Questions and
Answers Verified 100% Correct
The entrance to indoor locations containing positive pressure gases other than
oxygen and medical air shall be labeled including: ______________
A Positive Pressure Gases
B No Smoking or Open Flame
C Room May Have Insufficient Oxygen - Open Door and Allow Room to
Ventilate Before Entering
D All of the above - ANSWER -All of the above

If heat is necessary to keep an indoor manifold enclosure above 20oF and the
manufacturer's recommended minimum ambient temperature it may be provided
by:
A Any in-room electric heating element
B In-room heating element with a maximum temperature of 266oF
C Fuel fired equipment in the room
D Heat blanket with maximum temperature of 300oF - ANSWER -In-room
heating element with a maximum temperature of 266oF

Two entries/exits are required for ___________________ gas supplies.
A Outdoor Cryogenic Gas Supplies only
B Outdoor Cylinder Gas Supplies in an enclosure >200ft2 only
C Indoor Gas Supplies only
D Both A & B - ANSWER -Both A & B

An existing outdoor Nitrous Oxide Manifold may continued to be used since it
seldom falls below 20oF.
A TRUE
B FALSE - ANSWER -FALSE

Indoor manifold rooms must be separated by ______ fire rated walls and floors
with a ______ opening.
A 1 hour, 1 hour
B 2 hour, 1 hour
C 1 hour, 3/4 hour
D No wall fire rating is required - ANSWER -1 hour, 3/4 hour

, Only full or in-use cylinders are required to be prevented from falling by being
secured with racks, chains, or other fastenings. Empty cylinders are not required to
be secured.
A TRUE
B FALSE - ANSWER -FALSE

Oxygen supplies greater than 350 psig may not use _______________ materials.
A Copper
B Stainless Steel
C Polymeric
D Brass - ANSWER -Polymeric

Gas supply source relief valve vent lines must be labeled ____________.
A with the proper pressure in psig
B with medical gas pipe labels
C by painting the correct CGA color
D in a manner to distinguish them from medical gas pipeline - ANSWER -in a
manner to distinguish them from medical gas pipeline

Manifold rooms utilizing natural ventilation shall
A have 2 non-closeable louvered openings
B be vented directly to the outside without ductwork
C provide cross ventilation
D all of the above - ANSWER -all of the above

Indoor manifold enclosures utilizing mechanical ventilation shall maintain
positive pressure
A TRUE
B FALSE - ANSWER -FALSE

A gas cylinder header consists of all of the following components with the
exception of:
A Filter
B Pressure Indicator
C Local alarm
D Regulator - ANSWER -Local alarm

A header for cryogenic liquid containers must have a ________________.
A pressure relief valve
B purge ports

, C quick disconnect
D none of the above - ANSWER -pressure relief valve

A __________ must be installed at the connection between the header and each
cylinder lead (pigtail) to prevent a single pigtail leak from causing pressure loss of
all cylinders connected to that same header.
A Warning light
B Local Alarm
C Gauge
D Check Valve - ANSWER -Check Valve

Full and empty cylinder storage may not be in the same enclosure as a central
supply system (medical or support gas manifold).
A TRUE
B FALSE - ANSWER -FALSE

Permitted areas for piped medical gases outlets include all of the following with
exception of:
A Direct Respiration by a Patient
B Calibration of patient respiration devices
C Endoscopy or Sterile Processing Instrument Blow Down
D Simulation centers for education of health care professionals - ANSWER -
Endoscopy or Sterile Processing Instrument Blow Down

An auxiliary valve source connection must consist of all of the following except:
A Installed on the source side (upstream) of the source valve
B Must the same size as the main line
C Include a removable plug or cap
D Must be normally closed and secured - ANSWER -Installed on the source side
(upstream) of the source valve

Facilities may supply ______________ through a Concentrator Supply Source.
A USP 93 Oxygen
B 99% Oxygen Only
C Compressed Medical Air
D Nitrous Oxide - ANSWER -USP 93 Oxygen

The primary and secondary headers for a gas cylinder manifold must each have an
equal number of cylinders.
A TRUE

Geschreven voor

Instelling
NFPA 99 6040
Vak
NFPA 99 6040

Documentinformatie

Geüpload op
15 november 2025
Aantal pagina's
18
Geschreven in
2025/2026
Type
Tentamen (uitwerkingen)
Bevat
Vragen en antwoorden

Onderwerpen

$13.89
Krijg toegang tot het volledige document:

Verkeerd document? Gratis ruilen Binnen 14 dagen na aankoop en voor het downloaden kun je een ander document kiezen. Je kunt het bedrag gewoon opnieuw besteden.
Geschreven door studenten die geslaagd zijn
Direct beschikbaar na je betaling
Online lezen of als PDF

Maak kennis met de verkoper

Seller avatar
De reputatie van een verkoper is gebaseerd op het aantal documenten dat iemand tegen betaling verkocht heeft en de beoordelingen die voor die items ontvangen zijn. Er zijn drie niveau’s te onderscheiden: brons, zilver en goud. Hoe beter de reputatie, hoe meer de kwaliteit van zijn of haar werk te vertrouwen is.
TopGradeGuru Teachme2-tutor
Volgen Je moet ingelogd zijn om studenten of vakken te kunnen volgen
Verkocht
15
Lid sinds
1 jaar
Aantal volgers
0
Documenten
2399
Laatst verkocht
4 weken geleden
GRADEHUB

We provide access to a wide range of professionally curated exams for students and educators. It offers high-quality, up-to-date assessment materials tailored to various subjects and academic levels. With instant downloads and affordable pricing, it\'s the go-to resource for exam preparation and academic success.

1.5

2 beoordelingen

5
0
4
0
3
0
2
1
1
1

Recent door jou bekeken

Waarom studenten kiezen voor Stuvia

Gemaakt door medestudenten, geverifieerd door reviews

Kwaliteit die je kunt vertrouwen: geschreven door studenten die slaagden en beoordeeld door anderen die dit document gebruikten.

Niet tevreden? Kies een ander document

Geen zorgen! Je kunt voor hetzelfde geld direct een ander document kiezen dat beter past bij wat je zoekt.

Betaal zoals je wilt, start meteen met leren

Geen abonnement, geen verplichtingen. Betaal zoals je gewend bent via iDeal of creditcard en download je PDF-document meteen.

Student with book image

“Gekocht, gedownload en geslaagd. Zo makkelijk kan het dus zijn.”

Alisha Student

Bezig met je bronvermelding?

Maak nauwkeurige citaten in APA, MLA en Harvard met onze gratis bronnengenerator.

Bezig met je bronvermelding?

Veelgestelde vragen