Oxygen monitors, when used for oxygen concentrator supply units,
shall be calibrated at what interval?
A At least once a day
B At least every three months
C At least yearly
D They do not need to be calibrated - ANSWERS-At least every three
months
Which of the following operation, management, and maintenance
items are enforceable for both new and existing category 1 facilities?
A Manifold Room Ventilation & Temperatures
B Medical Air Usage (only for human respiration)
C Equipment Maintenance, Labeling, & personnel qualifications
D All of the above - ANSWERS-All of the above
Under the Qualifications section for Maintenance Programs, the
systems must be maintained by someone who is technically
competent on the specific equipment installed at the facility and by
demonstrating which of the following?
A ASSE 6040 certification
B ASSE 6030 certification
C Training Certificate approved by the health care facility
D Any of the above - ANSWERS-Any of the above
,What test(s) shall be performed for nonstationary booms &
articulating assemblies with flexible connectors every 18 months or as
determined by a risk assessment.
A Joints examined for leakage
B Safe working condition of flexible assemblies (hoses)
C Leak test DISS connections
D All of the above - ANSWERS-All of the above
Headwalls utilizing flexible connectors shall be tested every 18
months.
A TRUE
B FALSE - ANSWERS-FALSE
Shutoff valves labels shall be updated when modifications are made
which change the areas served.
A TRUE
B FALSE - ANSWERS-TRUE
The bulk cryogenic liquid supply systems shall include these
inspections and/or documentation:
A vaporizer sizing criteria documentation for the facility by the
supplier
B annual review of sufficient capacity
C annual inspection by equipment owner
, D All of the above - ANSWERS-All of the above
Testing documents of Area, Master, and Local alarms for periodic
audible and visual alarm indicators shall be maintained
_____________.
A for 30 years
B for 15 years
C until the next test is performed
D records are not required to be kept - ANSWERS-until the next test
is performed
Medical-surgical vacuum station inlet preventative maintenance
requirements includes all of the below except:
A performance testing on a regular pm schedule determined by the
facility maintenance staff
B be flushed and disinfected annually
C must have at least 3.0 SCFM
D must not reduce the system's vacuum level below 12"Hg -
ANSWERS-be flushed and disinfected annually
Medical Air system carbon monoxide monitors must be calibrated at
least _________ or more often if recommended by the manufacturer.
A Annually
B Bi-Annually
C Quarterly
D Periodically - ANSWERS-Annually