CCDS exam review Questions and Correct Answers |
Latest Update
What lead insulation materials are known for reducing fibroblast
ingrowth and susceptibility to Metal-ion induced oxidation (MIO) and
Environmental Stress-Cracking (ESC)
Ans: ePTFE and 80A polyurethane
What is the most common cause of safety pacing or "ventricular safety
standby?"
Assignment Expert
Ans: Ventricular Ectopy
Guru01 - Stuvia
Ventricular safety standby is a feature designed to avoid the
inappropriate inhibition of V-pacing due to crosstalk oversensing of an
Ap event on the ventricular EGM. Not sure why the book says that PVCs
are the most common cause, perhaps because bipolar leads have made
2026
Crosstalk such a rare-occurring issue now.
Which electrode material is most likely to result in higher pacing
©
thresholds and higher degree of polarization?
Ans: Elgiloy
What phase of the action potential correlates to relative refractory
period?
Ans: 3
The ventriculoatrial conduction time of a pacemaker patient with a
typical endless-loop-tachycardia may be calculated by which equation?
Ans: (upper rate limit interval) - ( AV delay + PVARP) ????????????????
or is it just MTR (URL) - AVD
, 2 for specific request mail
For treatment of arrhythmias, most drugs work by altering _______ or
__________.
Ans: Conductivity or Rerfractoriness
1Hz = ??bpm
Ans: 60bpm
What would cause a more significant drain on battery charge... ATP
before charge or ATP during charge?
Ans: ATP DURING charge
Assignment Expert
Increased current density is expected with what type of lead feature?
Guru01 - Stuvia
Ans: Small tip (cathode) surface area
________ is the component of a device that converts both positive and
negative components into a signal with only one component.
2026
Ans: Rectifier
©
Which of the following would account for the most common cause of
inappropriate Mode switching?
A: Ventricular lead sensing atrial signals
B: Lead Insulation break
C: Oversensing EMI
D: Atrial lead sensing Ventricular Signals
Ans: D: Atrial lead sensing Ventricular Signals
________ ________ is a closed loop sensor used by Biotronik for sensor-
driven pacing rate.
Ans: Myocardial Contractility